PROXIMATE** LINEAR CUTTER RELOAD - STANDARD
Report
- Report Number
- 3005075853-2013-01565
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- March 8, 2013
- Report Date
- March 10, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE RELOAD WAS RECEIVED IN GOOD VISUAL CONDITION. THE CARTRIDGE RELOAD HAD THE SWING TAB IN THE LOCKED POSITION, AND THE PROXIMAL ONE DRIVER UP AND THE REMAINING DRIVERS WERE DOWN WITH STAPLES PRESENT. IT COULD NOT BE DETERMINED IF THE REPORTED INCIDENT WAS THE RESULT OF A PREMATURE LOCKOUT, OR THE RESULT OF AN INTERRUPTED FIRING STROKE. THE SWING TAB WAS RESET AND THE CARTRIDGE WAS LOADED INTO A TEST DEVICE FOR FUNCTIONAL TESTING. THE DEVICE ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER SHAPE. WHILE NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT, IT SHOULD BE NOTED THAT 100% INSPECTION IS PERFORMED BY MEANS OF AUTOMATED VISION SYSTEM TO INSURE THE SWING TAB IS PRESENT AND UNLOCKED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, NO STAPLES CAME OUT AFTER THE FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 137305 | PROXIMATE** LINEAR CUTTER RELOAD - STANDARD | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | J4C710 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |