FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3033258 · Received April 3, 2013

Report

Report Number
2531779-2013-03489
Event Type
Malfunction
Date Received
April 3, 2013
Report Date
March 6, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 06/18/2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: INVESTIGATORS CONFIRMED THE AUTO TIMEOUT FEATURE WAS SET TO 12 HOURS. THE LAST AUTO OFF TIMEOUT PUMP SUSPENDED¿ WAS RECORDED IN THE ALARM HISTORY. THE ALARM HISTORY SHOWS SEVERAL 150 ALARMS, HOWEVER, NONE OF THEM OCCUR WITHIN THE HALF AN HOUR AS MENTIONED IN THE SCRIPTS AND NONE ARE WITHIN A 12 HOUR TIME PERIOD. FOR THE INVESTIGATION THE PUMPS AUTO OFF TIMEOUT FEATURE WAS SET TO 1 HR; NO KEY PRESSES WERE EXECUTED WITHIN THE 1 HR TIME PERIOD. INVESTIGATORS WERE UNABLE TO DUPLICATE THE COMPLAINT OF AUTO OFF PRIOR TO THE PROGRAMMED DURATION PERIOD DURING THE INVESTIGATION.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (AUTO OFF) ISSUE. THE PUMP IS REPORTEDLY ENACTING THE AUTO OFF FEATURE SOONER THAT THE PATIENT EXPECTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
135788 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 33 YR