PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3033243 · Received April 3, 2013

Report

Report Number
2032227-2013-01248
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 22, 2013
Report Date
March 22, 2013
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP ALSO HAD A MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AND ALARM AND INSULIN SHOOTING OUT DURING THE MANUAL PRIME. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.