PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3033243
·
Received April 3, 2013
Report
- Report Number
- 2032227-2013-01248
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- March 22, 2013
- Report Date
- March 22, 2013
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP ALSO HAD A MISSING END CAP STICKER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED AND ALARM AND INSULIN SHOOTING OUT DURING THE MANUAL PRIME. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.