FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3033146 · Received April 3, 2013

Report

Report Number
1416980-2013-08242
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 13, 2013
Report Date
March 13, 2013
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING PREVENTATIVE MAINTENANCE. THE ROOT CAUSE OF THE DOOR OPEN / CLOSE CLAMP ALARM IS UNKNOWN. TO CORRECT THE CONDITION, THE DOOR LATCH MAGNET WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING THE PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A DOOR OPEN / CLOSE CLAMP ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136558 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1