FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3033115 · Received April 1, 2013

Report

Report Number
1720753-2013-04447
Event Type
Malfunction
Date Received
April 1, 2013
Date of Event
January 23, 2013
Report Date
April 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE FILAMENT DRIVER BOARD. THE SYSTEM OPERATES AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED A FILAMENT REGULATOR ERROR DURING A PROCEDURE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132237 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1