ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-01564
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- December 26, 2012
- Report Date
- December 28, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE A WAS RETURNED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS FUNCTIONALLY TESTED ON THE GENERATOR AND THE DEVICE ACTIVATED WITH BOTH MAX AND MIN BUTTON. THE DEVICE WAS DISASSEMBLED TO INSPECT INTERNAL COMPONENTS. CORROSION WAS FOUND AT THE HAND ACTIVATION DOMES. IT IS PROBABLE THAT THIS MAY HAVE AFFECTED THE FUNCTIONALITY OF THE HAND ACTIVATION BUTTONS. DEVICE B WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED AND THE BLADE COMPLETELY BROKE OFF DURING ANALYSIS. DEVICE C WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED - EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICES WERE FUNCTIONALLY TESTED WITH A GENERATOR. BOTH DEVICES ACTIVATED WITH BOTH MAX AND MIN BUTTONS. DURING FUNCTIONAL TESTING ON THE GEN11 GENERATOR, THE 'INSTRUMENT ERROR' ALERT WAS DISPLAYED; AND WHEN TESTED ON THE GEN04 GENERATOR, AN ERROR CODE 5 (INSTRUMENT ERROR) WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICES TO STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 OR INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN YELLOW ALERT SCREENS, SUCH AS 'REPLACE INSTRUMENT' WITH THE GEN11 LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE. DEVICE B ADDITIONAL INFORMATION: BATCH # J9233M, EXPIRATION DATE: 7/24/2017, MANUFACTURING DATE: 8/24/2012. DEVICE C ADDITIONAL INFORMATION: BATCH # G9KD1Z; EXPIRATION DATE: 5/16/2015; MANUFACTURING DATE: 6/16/2010.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC ASSISTED DISTAL GASTRECTOMY PROCEDURE, HAND ACTIVATION BUTTON WAS NOT WORKING PROPERLY. IT WORKED INTERMITTENTLY. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE. PROCEDURE WAS PROLONGED 20 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136342 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR, HANDPIECE |