FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3033090 · Received April 1, 2013

Report

Report Number
1720753-2013-04456
Event Type
Malfunction
Date Received
April 1, 2013
Date of Event
March 22, 2013
Report Date
April 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE PLUG WAS REASSEMBLED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE FIELD ENGINEER REPORTED THAT THE INTERCONNECT PLUG WAS DISMANTLED AND THE SYSTEM WAS UNUSABLE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132983 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1