FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 3033080 · Received April 1, 2013

Report

Report Number
1720753-2013-04466
Event Type
Malfunction
Date Received
April 1, 2013
Date of Event
March 14, 2013
Report Date
April 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE RESISTANCE ON THE IRIS POTENTIOMETER WAS ADJUSTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM KEPT DISPLAYING A "BAD IRIS ERROR." ON THE 9600, THIS ERROR MESSAGE WILL PREVENT THE X-RAY FUNCTION FROM WORKING AND SHUT THE SYSTEM DOWN. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.