FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 3033053 · Received April 1, 2013

Report

Report Number
1720753-2013-04485
Event Type
Malfunction
Date Received
April 1, 2013
Date of Event
March 21, 2013
Report Date
April 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE LEG EXTENSION SWITCH WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM HAD AN INTERMITTENT ACCESSORY ERROR MESSAGE. THIS ERROR WILL PREVENT THE SYSTEM FROM BOOTING, RESULTING IN A RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132811 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1