FDA Adverse Event Malfunction Summary report: N

7700

MDR report key: 3033049 · Received April 1, 2013

Report

Report Number
9680959-2013-00692
Event Type
Malfunction
Date Received
April 1, 2013
Date of Event
March 15, 2013
Report Date
April 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT NO IMAGE WERE DISPLAYED WHEN PERFORMING FLUOROSCOPY. THIS ISSUE WOULD PREVENT THE LIVE IMAGE FROM BEING VIEWED, THEREBY MAKING THE SYSTEM UNUSABLE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132114 7700 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7700

Patients

Seq Age Sex Outcome Treatment
1