FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3033010 · Received April 3, 2013

Report

Report Number
1416980-2013-07401
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
January 9, 2013
Report Date
March 5, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. HOMECHOICE APD SYSTEMS TRAINER'S GUIDE. SECTION 12 "PROGRAMMING A PRESCRIPTION" IN 12.3.3 ON PAGE 12-6 GIVES THE WARNING THAT "IF YOU SET THE MINIMUM DRAIN VOLUME% TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD CAUSE AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 20:35:50. DURING NIGHT DRAIN CYCLE EIGHT, THE PATIENT'S ULTRAFILTRATION READING WAS 1665ML, INDICATING THE HOME PATIENT (HP) DRAINED 1665ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2200ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136748 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1