FDA Adverse Event Malfunction Summary report: N

EMERGE?

MDR report key: 3032843 · Received April 3, 2013

Report

Report Number
2134265-2013-01884
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 5, 2013
Report Date
March 5, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
K113220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE EMERGE CATHETER WAS RECEIVED WITH NO ORIGINAL PACKAGING OR OTHER DEVICES. THERE WAS BLOOD IN THE BALLOON AND INFLATION LUMEN. MAGNIFIED INSPECTION REVEALED NO DAMAGE. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE ALIGNED WITH THE DISTAL EDGE OF THE MARKERBAND. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE CALCIFIED AND MODERATELY TORTUOUS RIGHT CORONARY ARTERY. THE 1.20MM X 8MM EMERGE BALLOON WAS INFLATED FOUR TIME (INFLATIONS 1-3 UNKNOWN ATMS) AND RUPTURED AT 14ATMS ON THE FOURTH INFLATION. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE CALCIFIED AND MODERATELY TORTUOUS RIGHT CORONARY ARTERY. THE 1.20MM X 8MM EMERGE BALLOON WAS INFLATED FOUR TIME (INFLATIONS 1-3 UNKNOWN ATMS) AND RUPTURED AT 14ATMS ON THE FOURTH INFLATION. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
137244 EMERGE? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493919308120 15713289

Patients

Seq Age Sex Outcome Treatment
1