GREENLIGHT HPS BPH FIBER OPTIC
Report
- Report Number
- 2937094-2013-00379
- Event Type
- Malfunction
- Date Received
- March 29, 2013
- Date of Event
- December 6, 2012
- Report Date
- December 6, 2012
- Manufacturer
- AMS INNOVATIVE CENTER-SAN JOSE
- Product Code
- GEX
- PMA / PMN Number
- K062719
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE COMPONENT CODES FOR FIBER AND CAP REFER TO THE RESULTS CODE FOR THERMAL PROBLEM. FIBER ANALYSIS: THE FIBERS GLASS CAP WAS FOUND TO BE DETACHED; THE FIBER BROKEN PROXIMAL TO THE FIBER/CAP GLUE ZONE. THE FIBER CAP WAS NOT RETURNED BY THE CUSTOMER. THERE WAS NO INDICATION OR REPORT OF ANY PART OF THE DEVICE REMAINING INSIDE THE PATIENT. THE FIBER/ CAP CONDITIONS WOULD RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE THAT CONTRIBUTED TO THIS FAILURE WAS SUSPECTED TO BE RELATED TO HEAT ACCUMULATION DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE. CAP WEAR WAS ACCELERATED LIMITING THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBER TIP WAS DAMAGED. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.