FDA Adverse Event Malfunction Summary report: N

GREENLIGHT HPS BPH FIBER OPTIC

MDR report key: 3032786 · Received March 29, 2013

Report

Report Number
2937094-2013-00379
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
December 6, 2012
Report Date
December 6, 2012
Manufacturer
AMS INNOVATIVE CENTER-SAN JOSE
Product Code
GEX
PMA / PMN Number
K062719
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPONENT CODES FOR FIBER AND CAP REFER TO THE RESULTS CODE FOR THERMAL PROBLEM. FIBER ANALYSIS: THE FIBERS GLASS CAP WAS FOUND TO BE DETACHED; THE FIBER BROKEN PROXIMAL TO THE FIBER/CAP GLUE ZONE. THE FIBER CAP WAS NOT RETURNED BY THE CUSTOMER. THERE WAS NO INDICATION OR REPORT OF ANY PART OF THE DEVICE REMAINING INSIDE THE PATIENT. THE FIBER/ CAP CONDITIONS WOULD RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE THAT CONTRIBUTED TO THIS FAILURE WAS SUSPECTED TO BE RELATED TO HEAT ACCUMULATION DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE. CAP WEAR WAS ACCELERATED LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBER TIP WAS DAMAGED. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
131162 GREENLIGHT HPS BPH FIBER OPTIC POWERED SURGICAL LASER INSTRUMENT GEX AMS INNOVATIVE CENTER-SAN JOSE 10-2090 219H

Patients

Seq Age Sex Outcome Treatment
1