FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

MDR report key: 3032783 · Received March 29, 2013

Report

Report Number
1419937-2013-00175
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
MEDELA, INC
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SOURCE WAS SENT TO THE CUSTOMER. ATTEMPTS TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL. THE PRODUCT INVOLVED IN THE COMPLAINT HAS NOT BEEN RECEIVED FOR TESTING/ANALYSIS. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. HOWEVER, THE CUSTOMER REPORTED THAT THE HOUSING COMPLETELY FRACTURED APART SO THAT THE SCREWS CAN NOT HOLD THE HOUSING TOGETHER, EXPOSING THE WIRES, WHICH IS A SAFETY RISK. THIS ISSUE WITH A DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMER TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. SHOULD ADDITIONAL INFORMATION OR THE ORIGINAL PRODUCT BE RECEIVED, RESULTING IN NEW, CHANGED, OR CORRECT INFORMATION, A FOLLOW UP REPORT WILL BE FILED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT THE HOUSING ON THE TRANSFORMER CAME APART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
131161 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC 57063/9207010 REV L-2011

Patients

Seq Age Sex Outcome Treatment
1