NUTRILINE 2D PICC
Report
- Report Number
- 2245270-2013-00016
- Event Type
- Malfunction
- Date Received
- March 29, 2013
- Date of Event
- February 28, 2013
- Report Date
- March 28, 2013
- Manufacturer
- VYGON GMBH
- Product Code
- LJS
- PMA / PMN Number
- K052475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NURSE
Narratives
ALTHOUGH, THE FAILED SAMPLE WAS NOT RETURNED TO VYGON US, THE COMPLAINT HAS BEEN FORWARDED TO VYGON (B)(4) (THE MANUFACTURER) FOR EVALUATION AND INVESTIGATION. THE RESULTS OF THIS INVESTIGATION WILL BE FORWARDED IN A FOLLOW-UP MDR WITHIN 30 OF ITS CONCLUSION.
A NUTRILINE PICC WAS PLACED AND ENCOUNTER A PRODUCT PROBLEM FIVE DAYS LATER. WHILE THE TEGADERM WAS BEING REMOVED DURING A DRESSING CHANGE, THE CATHETER SNAPPED AT THE PLACE WHERE IT CONNECTS TO THE WINGED HUB. THERE WAS NO UNUSUAL FORCE WITH REMOVING THE TEGADERM. IT WAS THEN NOTED THAT ABOUT 1CM ABOVE THE INSERTION SITE THERE WAS A FIBROUS 2CM IN LENGTH AREA FOLLOWING THE PATH OF THE CATHETER. THE CATHETER WOULD NOT PULL BACK AND IT TOOK OVER 30 MINUTES TO REMOVE THE CATHETER. IT WOULD PULL BACK IN SMALL AMOUNTS WHILE LEG WAS BEING MASSAGED. THE DOCTOR HAS TO ASSIST WITH THE REMOVAL OF THIS CATHETER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 129863 | NUTRILINE 2D PICC | NEONATAL PICC | LJS | VYGON GMBH | 1252.30G | 141112GA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |