MUSTANG¿
Report
- Report Number
- 2134265-2013-01977
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- March 7, 2013
- Report Date
- March 7, 2013
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- LIT
- PMA / PMN Number
- K103751
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION WAS LOCATED IN A MODERATELY TORTUOUS AND CALCIFIED FOREARM SHUNT ANASTOMOSIS AREA. ON THE FIRST INFLATION THE MUSTANG 6.0 X 40, 40CM WAS INFLATED TO TWENTY FOUR ATMOSPHERES. ON THE SECOND INFLATION THE BALLOON WAS INFLATED TO TWENTY FOUR ATMOSPHERES AND RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 135817 | MUSTANG¿ | CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL | LIT | BOSTON SCIENTIFIC - GALWAY | H74939171060440 | 0015667815 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INTRODUCER SHEATH: MOSQUITO| GUIDEWIRE-RADIFOCUS |