FDA Adverse Event Other Summary report: N

DISPOSABLE FENESTRATED GRASPER TIP

MDR report key: 3032578 · Received March 29, 2013

Report

Report Number
1223422-2013-00012
Event Type
Other
Date Received
March 29, 2013
Report Date
March 29, 2013
Manufacturer
MICROLINE SURGICAL INC
Product Code
GEI
PMA / PMN Number
K981389
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ONE REF 3222 WITH AN UNKNOWN LOT NUMBER HAVING A BROKEN HINGE WAS RETURNED. THE HINGE OF THE BOTTOM JAW WAS BROKEN. THE FRACTURE OF THE BROKEN HINGE APPEARED TO BE A DUCTILE FRACTURE, WHICH IS CONSISTENT WITH APPLYING EXCESSIVE FORCE.

Description of Event or Problem · 1

THE HINGE BROKE INSIDE PT AND WAS RETRIEVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
130038 DISPOSABLE FENESTRATED GRASPER TIP DISPOSABLE FENESTRATED TIP GEI MICROLINE SURGICAL INC 3222 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1