FDA Adverse Event
Other
Summary report: N
DISPOSABLE FENESTRATED GRASPER TIP
MDR report key: 3032578
·
Received March 29, 2013
Report
- Report Number
- 1223422-2013-00012
- Event Type
- Other
- Date Received
- March 29, 2013
- Report Date
- March 29, 2013
- Manufacturer
- MICROLINE SURGICAL INC
- Product Code
- GEI
- PMA / PMN Number
- K981389
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ONE REF 3222 WITH AN UNKNOWN LOT NUMBER HAVING A BROKEN HINGE WAS RETURNED. THE HINGE OF THE BOTTOM JAW WAS BROKEN. THE FRACTURE OF THE BROKEN HINGE APPEARED TO BE A DUCTILE FRACTURE, WHICH IS CONSISTENT WITH APPLYING EXCESSIVE FORCE.
Description of Event or Problem · 1
THE HINGE BROKE INSIDE PT AND WAS RETRIEVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 130038 | DISPOSABLE FENESTRATED GRASPER TIP | DISPOSABLE FENESTRATED TIP | GEI | MICROLINE SURGICAL INC | 3222 | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |