FDA Adverse Event Other Summary report: N

M/L-10 CLIP APPLIER HANDPIECES REUSABLE

MDR report key: 3032577 · Received March 29, 2013

Report

Report Number
1223422-2013-00011
Event Type
Other
Date Received
March 29, 2013
Report Date
March 28, 2013
Manufacturer
MICROLINE SURGICAL INC.
Product Code
GCJ
PMA / PMN Number
K013695
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ONE REF 1002 M/L-10 CLIP APPLIER WAS RETURNED, DECONTAMINATED AND INVESTIGATED. THE RETURNED CLIP APPLIER WAS RECEIVED WITH THE JAWS APART AND TRIGGER IN THE HOME POSITION. THE TRIGGER OPERATED FREELY AND THE DEVICE APPEARED TO BE IN GOOD MECHANICAL CONDITION. THE JAWS WERE MEASURED AND FOUND TO OUT OF SPECIFICATION. THE DAMAGED FOUND IS CONSISTENT WITH MISHANDLING. ALTHOUGH DAMAGE WAS FOUND, THE ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

LOOSE CLIPS WERE FOUND NOT CLOSED IN THE ABDOMEN OF THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
130356 M/L-10 CLIP APPLIER HANDPIECES REUSABLE M/L-10 CLIP APPLIER GCJ MICROLINE SURGICAL INC. 1002 00105133

Patients

Seq Age Sex Outcome Treatment
1