FDA Adverse Event Other Summary report: N

EON

MDR report key: 3032377 · Received March 29, 2013

Report

Report Number
1627487-2013-06107
Event Type
Other
Date Received
March 29, 2013
Date of Event
March 5, 2012
Report Date
April 7, 2016
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE
Health Professional
N

Narratives

Additional Manufacturer Narrative · 1

THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 2. REF MFR REPORT: 1627487-2013-06108. IT WAS REPORTED THE PT HAS BEEN EXPERIENCING HEATING WHILE CHARGING. THE CHARGING SYSTEM HAS HEATED UP TO THE POINT THAT IT CAUSED A BLISTER. THE PT STATED HE HAS BEEN ATTEMPTING TO CONTACT SOMEONE ABOUT THE ISSUE FOR ABOUT A YR AND HAS MET WITH A SJM REP BUT CANNOT RECALL WHEN. A NEW LOW ENERGY CHARGER WAS SENT TO THE PT TO ADDRESS THE ISSUE. F/U INDICATES THE NEW CHARGER HAS RESOLVED THE HEATING ISSUE AND THE PT'S BLISTER IS HEALING WELL. ON (B)(4) 2012 ST JUDE MEDICAL, NEUROMODULATION DIV., SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS ARE EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
130371 EON SCS IPG LGW ST. JUDE MEDICAL - NEUROMODULATION 3716 171756

Patients

Seq Age Sex Outcome Treatment
1 41 YR Male Required Intervention IMPLANT DATE:| IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1194| SCS LEAD: MODEL 3210| SCS LEAD: MODEL 3245| IMPLANT DATE:| SCS ANCHOR: MODEL 1194| IMPLANT DATE:| IMPLANT DATE:| SCS LEAD: MODEL 3245| SCS LEAD: MODEL 3210