EON
Report
- Report Number
- 1627487-2013-06107
- Event Type
- Other
- Date Received
- March 29, 2013
- Date of Event
- March 5, 2012
- Report Date
- April 7, 2016
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
- Health Professional
- N
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REF MFR REPORT: 1627487-2013-06108. IT WAS REPORTED THE PT HAS BEEN EXPERIENCING HEATING WHILE CHARGING. THE CHARGING SYSTEM HAS HEATED UP TO THE POINT THAT IT CAUSED A BLISTER. THE PT STATED HE HAS BEEN ATTEMPTING TO CONTACT SOMEONE ABOUT THE ISSUE FOR ABOUT A YR AND HAS MET WITH A SJM REP BUT CANNOT RECALL WHEN. A NEW LOW ENERGY CHARGER WAS SENT TO THE PT TO ADDRESS THE ISSUE. F/U INDICATES THE NEW CHARGER HAS RESOLVED THE HEATING ISSUE AND THE PT'S BLISTER IS HEALING WELL. ON (B)(4) 2012 ST JUDE MEDICAL, NEUROMODULATION DIV., SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS ARE EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 130371 | EON | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3716 | 171756 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Male | Required Intervention | IMPLANT DATE:| IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1194| SCS LEAD: MODEL 3210| SCS LEAD: MODEL 3245| IMPLANT DATE:| SCS ANCHOR: MODEL 1194| IMPLANT DATE:| IMPLANT DATE:| SCS LEAD: MODEL 3245| SCS LEAD: MODEL 3210 |