EON MINI
Report
- Report Number
- 1627487-2013-02437
- Event Type
- Injury
- Date Received
- March 29, 2013
- Date of Event
- March 5, 2013
- Report Date
- March 5, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CORRECTION/REMOVAL NUMBERS: 1627487-05242011-002-R; 1627487-12192011-003-R. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION AND FIELD ADVISORIES. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-02438. IT WAS REPORTED THE PT EXPERIENCED POCKET HEATING AND BURNING AT HER IPG SITE WHICH ALSO OCCURS DURING CHARGING. THE PT STATED SHE WOULD LIKE TO HAVE HER IPG REPLACED. NO FURTHER INFO IS AVAILABLE AT THIS TIME. ON 08/01/2012, ST. JUDE MEDICAL NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 129786 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3058408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Required Intervention | SCS LEAD: MODEL 3219| IMPLANT DATE: |