EON MINI
Report
- Report Number
- 1627487-2013-01365
- Event Type
- Injury
- Date Received
- March 29, 2013
- Date of Event
- March 7, 2013
- Report Date
- March 7, 2013
- Manufacturer
- ST. JUDE MEDICAL- NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A 1627487-05242011-002-R, 1627487-07262012-002-R. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORY AND A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-01366. IT WAS REPORTED THE PATIENT IS EXPERIENCING POCKET HEATING WHILE CHARGING HIS IPG. A NEW LE CHARGER WAS SENT TO THE PATIENT TO ADDRESS THE ISSUE. FOLLOW UP INFORMATION IDENTIFIED THE REPLACEMENT CHARGING SYSTEM IS FUNCTIONING PROPERLY BUT HE IS STILL EXPERIENCING SOME HEATING WITH THE IPG ITSELF. THE PATIENT WAS ADVISED TO USE THE REPLACEMENT CHARGER FOR A FEW WEEKS TO SEE IF THE HEATING STOPS. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 130529 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL- NEUROMODULATION | 3788 | 2796639 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Required Intervention | SCS LEAD: MODEL 3219| IMPLANT DATE: |