FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3032295 · Received March 29, 2013

Report

Report Number
1627487-2013-01366
Event Type
Injury
Date Received
March 29, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-09262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULT - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-01365. IT WAS REPORTED THE PATIENT IS EXPERIENCING POCKET HEATING WHILE CHARGING HIS IPG. A NEW LE CHARGER WAS SENT TO THE PATIENT TO ADDRESS THE ISSUE. FOLLOW UP INFORMATION IDENTIFIED THE REPLACEMENT CHARGING SYSTEM IS FUNCTIONING PROPERLY BUT HE IS STILL EXPERIENCING SOME HEATING WITH THE IPG ITSELF. THE PATIENT WAS ADVISED TO USE THE REPLACEMENT CHARGER FOR A FEW WEEKS TO SEE IF THE HEATING STOPS. ON (B)(4) 2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
131210 CHARGING SYSTEM SCS CHARGER LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 2756233

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention IMPLANT DATE:| SCS LEAD: MODEL 3219