FDA Adverse Event Injury Summary report: N

EON MINI

MDR report key: 3032265 · Received March 29, 2013

Report

Report Number
1627487-2013-01357
Event Type
Injury
Date Received
March 29, 2013
Date of Event
March 6, 2013
Report Date
March 6, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2013-1356. THE PT HAS TWO LEADS IMPLANTED WITH THE SAME LOT NUMBER. IT WAS REPORTED THE PT SUFFERED A FALL AND LOST STIMULATION. X-RAYS WERE TAKEN AND SHOWED THE LEAD WAS PARTIALLY OUT OF THE IPG HEADER. THE PT UNDERWENT REVISION SURGERY ON (B)(6) 2013 WHERE THE LEAD WAS RE-INSERTED INTO THE IPG. THE PT IS NOW RECEIVING EFFECTIVE STIMULATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
131186 EON MINI SCS IPG LGW ST. JUDE MEDICAL - NEUROMODULATION 3788 3875424

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2)