FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 3032064 · Received March 6, 2013

Report

Report Number
3004464228-2013-00185
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
January 27, 2013
Report Date
February 4, 2013
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVAL. WE ARE UNABLE TO CONFIRM THE REPORTED MALFUNCTION OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO REPORTED HYPERGLYCEMIA. QUALIFICATION RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER GUIDE WARNS "CHECK THE INFUSION SITE AFTER INSERTION TO ENSURE THAT THE CANNULA WAS PROPERLY INSERTED. IT IS ALSO A GOOD IDEA TO CHECK YOUR BLOOD GLUCOSE ABOUT TWO HOURS AFTER EACH POD CHANGE AND TO CHECK THE INFUSION SITE PERIODICALLY," AND "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER". IT ADVISES "IF YOUR BLOOD GLUCOSE IS 250 MG/DL OR ABOVE, CHECK FOR KETONES. IF KETONES ARE NOT PRESENT, TAKE A CORRECTION BOLUS AS PRESCRIBED BY YOUR HEALTHCARE PROVIDER. CHECK BLOOD GLUCOSE AGAIN AFTER 2 HOURS. IF BLOOD GLUCOSE LEVELS HAVE NOT DECREASED, TAKE A SECOND BOLUS BY INJECTION, USING A STERILE SYRINGE. IF BLOOD GLUCOSE REMAINS HIGH AFTER ANOTHER 2 HOURS (A TOTAL OF 4 HOURS), REPLACE THE POD."

Description of Event or Problem · 1

THE CUSTOMER ACTIVATED THE POD AT 1:30 PM ON (B)(6). HER BLOOD GLUCOSE, CARBOHYDRATE INTAKE AND INSULIN TREATMENT FOR THE TWELVE HOURS OF WEAR. THE POD WAS DEACTIVATED AT 1:12 AM AND SHE COULD STILL SEE THE NEEDLE TIP IN THE CANNULA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96541 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L31010

Patients

Seq Age Sex Outcome Treatment
1 27 YR