EON MINI
Report
- Report Number
- 1627487-2013-12400
- Event Type
- Injury
- Date Received
- March 26, 2013
- Date of Event
- March 1, 2013
- Report Date
- March 1, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN.
DEVICE 1 OF 2. REFERENCE MFR REPORT #1627487-2013-12401. IT WAS REPORTED, THE PATIENT EXPERIENCES HEATING AT THE IPG POCKET SITE WITH STIMULATION ON, AND THE POCKET SITE GETS WARMER WHEN CHARGING. THE AREA IS ALSO SWOLLEN. A NEW CHARGER WAS SENT TO THE PATIENT. REPORTEDLY, THE PHYSICIAN PLANS SURGICAL INTERVENTION TO ADDRESS THE ISSUE. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 122966 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3553141 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Required Intervention | SCS ANCHORS: MODEL 1192 (3)| IMPLANT DATE:| SCS EXTENSION: MODEL 3341| IMPLANT DATE:| IMPLANT DATE:| IMPLANT DATE:| SCS LEAD: MODEL 3166| SCS LEAD: MODEL 3163| SCS LEAD: MODEL 3244| IMPLANT DATE: |