CD HORIZON® SPINAL SYSTEM
Report
- Report Number
- 1030489-2013-00923
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- March 5, 2013
- Report Date
- April 3, 2013
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HXX
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4): THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION; THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THERE IS NO DAMAGE NOTED TO MAS HEAD THREADED INTERFACE. VISUALLY CONFIRMED APPROXIMATELY ~3MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. DIMENSIONAL INSPECTION OF THE RELEVANT DIMENSION CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD, WITH THE TIP JUST BELOW THE FRACTURE SURFACES PLASTICALLY DEFORMED. ADDITIONALLY, THE ASSEMBLY ATTACHMENT PIN TO THE INTERNAL BUSHING HAS BEEN BROKEN, CONSISTENT WITH TORSIONAL OVERLOAD CONDITION. THE ABOVE FINDINGS ARE CONSISTENT WITH TORSIONAL OVERLOAD.
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN ANTERIOR CERVICAL DISCECTOMY AND FUSION. IT WAS REPORTED THAT WHILE TRYING TO BACK OUT A SCREW, THE SCREWDRIVER TIP BROKE AND REMAINED LODGED IN THE HEAD OF THE SCREW. THE SCREW WAS REMOVED AND REPLACED WITH ANOTHER SCREW. NO PATIENT COMPLICATIONS WERE REPORTED.