FDA Adverse Event Malfunction Summary report: N

CD HORIZON® SPINAL SYSTEM

MDR report key: 3031909 · Received April 3, 2013

Report

Report Number
1030489-2013-00923
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 5, 2013
Report Date
April 3, 2013
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HXX
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION; THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THERE IS NO DAMAGE NOTED TO MAS HEAD THREADED INTERFACE. VISUALLY CONFIRMED APPROXIMATELY ~3MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. DIMENSIONAL INSPECTION OF THE RELEVANT DIMENSION CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD, WITH THE TIP JUST BELOW THE FRACTURE SURFACES PLASTICALLY DEFORMED. ADDITIONALLY, THE ASSEMBLY ATTACHMENT PIN TO THE INTERNAL BUSHING HAS BEEN BROKEN, CONSISTENT WITH TORSIONAL OVERLOAD CONDITION. THE ABOVE FINDINGS ARE CONSISTENT WITH TORSIONAL OVERLOAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT AN ANTERIOR CERVICAL DISCECTOMY AND FUSION. IT WAS REPORTED THAT WHILE TRYING TO BACK OUT A SCREW, THE SCREWDRIVER TIP BROKE AND REMAINED LODGED IN THE HEAD OF THE SCREW. THE SCREW WAS REMOVED AND REPLACED WITH ANOTHER SCREW. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136111 CD HORIZON® SPINAL SYSTEM SCREWDRIVER HXX MEDTRONIC SOFAMOR DANEK USA, INC NA RS10M015

Patients

Seq Age Sex Outcome Treatment
1