UNICEL® DXI 800 ACCESS® IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2013-00337
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- February 20, 2013
- Report Date
- March 4, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- DHA
- PMA / PMN Number
- K023764
- Removal / Correction Number
- Z-1736-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) NOTED THE SAMPLE PROBE CARRY-OVER TEST FAILED AND INSTALLED A NEW SAMPLE PROBE AND CARRY-OVER TEST PASSED. THE FSE COMPLETED HIGH SENSITIVITY SYSTEM CHECK WITH PASSING RESULTS. QUALITY CONTROL (QC) WAS WITHIN THE LABORATORY'S ESTABLISHED LIMITS. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. THE PATIENT'S SAMPLES WERE COLLECTED IN A 13X100 MM GREEN TOP TUBE AND CENTRIFUGED AT 3,000 RPM (ROTATIONS PER MINUTE) FOR SEVEN MINUTES. THE SAMPLES WERE ANALYZED ON THE DAY OF COLLECTION AND REFRIGERATED AFTER THE ANALYSIS. SAMPLE QUALITY APPEARED NORMAL. THE LABORATORY PERFORMS QUALITY CONTROL (QC) EVERY EIGHT HOURS. QC WAS WITHIN RANGE AT THE TIME OF THE EVENT. THE CUSTOMER REANALYZED RECENT AND ANY NEW POSITIVE RESULTS FOR ACCURACY VERIFICATION. THIS MEDWATCH REPORT IS RELATED TO MDR 2122870-2013-00338.
THE CUSTOMER REPORTED ERRONEOUS BETA HUMAN CHORIONIC GONADOTROPHIN (HCG) RESULTS, FOR TWO PATIENTS, ON TWO SEPARATE DAYS, INVOLVING THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM UTILIZED WITH THE ACCESS® TOTAL SSHCG ASSAY AND CALIBRATORS. SUBSEQUENT ANALYSES OF THE PATIENTS' SAMPLES, ON AN ALTERNATE UNICEL DXI 800 SYSTEM, RECOVERED LOWER RESULTS. A SECOND SAMPLE WAS COLLECTED EIGHT HOURS AFTER AND ALSO PRODUCED A LOWER RESULT. THE ERRONEOUS RESULTS WERE RELEASED OUT OF THE LABORATORY AND QUESTIONED BY THE PHYSICIAN. THE CUSTOMER COULD NOT CONFIRM IF THERE WAS ANY CHANGE TO PATIENT CARE. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE SAMPLE WAS INITIALLY TESTED FROM THE PRIMARY TUBE ON THE AUTOMATION LINE AND THE REANALYSIS WAS MANUALLY LOADED THROUGH THE SPECIMEN PROCESSING UNIT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THIS REPORT REFERENCES PATIENT ONE OF TWO FOR THE EVENT DATE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 134562 | UNICEL® DXI 800 ACCESS® IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | DHA | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |