FDA Adverse Event Death Summary report: N

DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

MDR report key: 3031568 · Received April 2, 2013

Report

Report Number
2649622-2013-02590
Event Type
Death
Date Received
April 2, 2013
Date of Event
March 2, 2013
Report Date
May 22, 2013
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCTS: PRODUCT ID D354TRG, SERIAL# (B)(4), IMPLANTED: (B)(6) 2013, PRODUCT ID 693565, SERIAL# (B)(4), IMPLANTED: (B)(6) 2013, PRODUCT ID 407652, SERIAL# (B)(4), IMPLANTED: (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT, VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)  SYSTEM AND THE NEXT DAY EXPIRED IN THE HOSPITAL DUE TO VENTRICULAR TACHYCARDIA (VT) AND VENTRICULAR FIBRILLATION (VF). THE STRIPS FROM THE MONITOR SHOWED A SLOW VENTRICULAR TACHYCARDIA (VT) UNDER THE DETECTION ZONE. THE CONCERN IS ABOUT WHY THE PATIENT DID NOT GOT SHOCKED FOR THE VENTRICULAR FIBRILLATION (VF) AND WHY PACING DID NOT HAPPEN. THE CUSTOMER ALLEGES THAT A POSSIBLE DEVICE MALFUNCTION LED TO THE DEATH OF THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134678 DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 439688

Patients

Seq Age Sex Outcome Treatment
1 00068 YR Death