ENDURANT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 3031565 · Received April 2, 2013

Report

Report Number
2953200-2013-00593
Event Type
Injury
Date Received
April 2, 2013
Date of Event
May 11, 2011
Report Date
May 14, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: (FILMS).

Additional Manufacturer Narrative · 1

(B)(4) METHOD: (FILM). RESULTS: INHERENT RISK OF PROCEDURE (ENDOLEAK); PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (TYPE II ENDOLEAK, MORPHOLOGY CHANGES). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (TYPE II ENDOLEAK, MORPHOLOGY CHANGES).

Description of Event or Problem · 1

IT WAS REPORTED THAT 2 YEAR FOLLOW-UP IMAGING REVEALED A SEPARATION BETWEEN THE IPSILATERAL LIMB 162495 AND THE BIFURCATED STENT GRAFT IN THE COMMON ILIAC ARTERY. THE PHYSICIAN BELIEVES A PERSISTENT TYPE II ENDOLEAK CAUSED A CHANGE IN THE PRESSURE DYNAMICS OF THE ANEURYSM SAC WHICH CAUSED THE STENT GRAFTS TO SEPARATE. AN ENDURANT 202080 WAS IMPLANTED SIX DAYS LATER COVERING THE SEPARATION AND RESOLVING THE TYPE III ENDOLEAK. TWO X-RAY IMAGES WERE REVIEWED AND SHOWED THE IPSILATERAL EXTENSION WAS OVERLAPPING THE BIFURCATE IPSILATERAL LIMB BY APPROXIMATELY 2-3 STENT RINGS (PER THE MARKERS). THE IPSILATERAL LIMB WAS RELATIVELY STRAIGHT ALONG ITS LENGTH. AN X-RAY FROM APPROXIMATELY SIX WEEKS AGO SHOWS THERE IS NO OVERLAPPING BETWEEN THE IPSILATERAL LIMB AND EXTENSION (MARKERS ARE NEARLY ALIGNED). THE ORIGIN OF THE IPSILATERAL LIMB IS MORE ANGULATED COMPARED TO THE PREVIOUS IMAGE. ASSESSMENT OF THE ENDOLEAK AND THE AMOUNT OF DISTAL SEALING IN THE RIGHT ILIAC COULD NOT BE PERFORMED. THE CAUSE OF THE SEPARATION IS UNKNOWN. IT IS POSSIBLE THAT ANEURYSM MORPHOLOGY CHANGES, IN COMBINATION WITH LACK OF DIAMETER OVERSIZING BETWEEN THE TWO LIMBS, MAY HAVE CONTRIBUTED TO THE SEPARATION.

Description of Event or Problem · 1

ADDITIONAL POST-IMPLANT FILMS WERE RECEIVED FROM THE 4-MONTHS FOLLOW-UP ON THE BIFURCATE WAS SEEN POSITIONED JUST BELOW THE RENAL ARTERIES. THERE IS A POSSIBLE PROXIMAL TYPE I ENDOLEAK. FOLLOWING THE IMPLANT OF A PALMAZ STENT IMPLANTED WITHIN THE ENDURANT AORTIC BODY, THE ENDOLEAK APPEARS TO HAVE RESOLVED. THE CAUSE OF THE PROXIMAL TYPE I ENDOLEAK IS UNKNOWN. THE IPSILATERAL LIMB IS ON THE RIGHT SIDE, AND AN EXTENSION IS SEEN OVERLAPPING THE LIMB BY APPROXIMATELY 3.5CM (PER THE MARKERS). THE IPSILATERAL LIMB IS ESSENTIALLY STRAIGHT IN THIS IMAGE. NO ENDOLEAK OR ANY STENT GRAFT ISSUES WERE OBSERVED ON FINAL ANGIOGRAM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134677 ENDURANT ABDOMINAL STENT GRAFT SYSTEM SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND V00389253

Patients

Seq Age Sex Outcome Treatment
1 00080 YR Required Intervention