ENDURANT ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2013-00593
- Event Type
- Injury
- Date Received
- April 2, 2013
- Date of Event
- May 11, 2011
- Report Date
- May 14, 2013
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
METHOD: (FILMS).
(B)(4) METHOD: (FILM). RESULTS: INHERENT RISK OF PROCEDURE (ENDOLEAK); PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (TYPE II ENDOLEAK, MORPHOLOGY CHANGES). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (TYPE II ENDOLEAK, MORPHOLOGY CHANGES).
IT WAS REPORTED THAT 2 YEAR FOLLOW-UP IMAGING REVEALED A SEPARATION BETWEEN THE IPSILATERAL LIMB 162495 AND THE BIFURCATED STENT GRAFT IN THE COMMON ILIAC ARTERY. THE PHYSICIAN BELIEVES A PERSISTENT TYPE II ENDOLEAK CAUSED A CHANGE IN THE PRESSURE DYNAMICS OF THE ANEURYSM SAC WHICH CAUSED THE STENT GRAFTS TO SEPARATE. AN ENDURANT 202080 WAS IMPLANTED SIX DAYS LATER COVERING THE SEPARATION AND RESOLVING THE TYPE III ENDOLEAK. TWO X-RAY IMAGES WERE REVIEWED AND SHOWED THE IPSILATERAL EXTENSION WAS OVERLAPPING THE BIFURCATE IPSILATERAL LIMB BY APPROXIMATELY 2-3 STENT RINGS (PER THE MARKERS). THE IPSILATERAL LIMB WAS RELATIVELY STRAIGHT ALONG ITS LENGTH. AN X-RAY FROM APPROXIMATELY SIX WEEKS AGO SHOWS THERE IS NO OVERLAPPING BETWEEN THE IPSILATERAL LIMB AND EXTENSION (MARKERS ARE NEARLY ALIGNED). THE ORIGIN OF THE IPSILATERAL LIMB IS MORE ANGULATED COMPARED TO THE PREVIOUS IMAGE. ASSESSMENT OF THE ENDOLEAK AND THE AMOUNT OF DISTAL SEALING IN THE RIGHT ILIAC COULD NOT BE PERFORMED. THE CAUSE OF THE SEPARATION IS UNKNOWN. IT IS POSSIBLE THAT ANEURYSM MORPHOLOGY CHANGES, IN COMBINATION WITH LACK OF DIAMETER OVERSIZING BETWEEN THE TWO LIMBS, MAY HAVE CONTRIBUTED TO THE SEPARATION.
ADDITIONAL POST-IMPLANT FILMS WERE RECEIVED FROM THE 4-MONTHS FOLLOW-UP ON THE BIFURCATE WAS SEEN POSITIONED JUST BELOW THE RENAL ARTERIES. THERE IS A POSSIBLE PROXIMAL TYPE I ENDOLEAK. FOLLOWING THE IMPLANT OF A PALMAZ STENT IMPLANTED WITHIN THE ENDURANT AORTIC BODY, THE ENDOLEAK APPEARS TO HAVE RESOLVED. THE CAUSE OF THE PROXIMAL TYPE I ENDOLEAK IS UNKNOWN. THE IPSILATERAL LIMB IS ON THE RIGHT SIDE, AND AN EXTENSION IS SEEN OVERLAPPING THE LIMB BY APPROXIMATELY 3.5CM (PER THE MARKERS). THE IPSILATERAL LIMB IS ESSENTIALLY STRAIGHT IN THIS IMAGE. NO ENDOLEAK OR ANY STENT GRAFT ISSUES WERE OBSERVED ON FINAL ANGIOGRAM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 134677 | ENDURANT ABDOMINAL STENT GRAFT SYSTEM | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V00389253 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00080 YR | Required Intervention |