FDA Adverse Event Malfunction Summary report: N

PROXIMATE** LINEAR CUTTER RELOAD - STANDARD

MDR report key: 3031555 · Received April 2, 2013

Report

Report Number
3005075853-2013-01556
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SWING TAB IN UNLOCKED POSITION. THE ANALYSIS RESULTS FOUND THAT THE CARTRIDGE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH NO INSTRUMENT. THE RELOAD WAS RECEIVED WITH THE 2 MOST PROXIMAL DRIVERS UP, AND THE SWING TAB IN THE UNLOCKED POSITION. IT SHOULD BE NOTED THAT AFTER THE STAPLE RETAINER HAS BEEN REMOVED, AN INADVERTENT MOVEMENT OF THE FIRING KNOB COULD CAUSE THE WEDGES TO MOVE FORWARD AND THEREFORE CAUSE THE STAPLES TO BECOME DISLODGED. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. THE CARTRIDGE WAS TESTED FOR FUNCTIONALITY WITH A TEST INSTRUMENT AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE INSTRUMENT FIRED WITHOUT ANY DIFFICULTIES AND THE STAPLES WERE NOTE TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN LAPAROSCOPY ASSISTED DISTAL GASTRECTOMY PROCEDURE THERE WAS AN INCOMPLETE STAPLE LINE. THE DEVICE WAS RELOADED AND THE CASE WAS COMPLETED WITH NO PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134639 PROXIMATE** LINEAR CUTTER RELOAD - STANDARD STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK J4AT33

Patients

Seq Age Sex Outcome Treatment
1