HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-08144
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- March 28, 2013
- Report Date
- March 28, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). PER BAXTER LABELING, USERS ARE INSTRUCTED THAT CLAMPS ON ANY UNUSED SOLUTION LINES MUST REMAIN CLOSED WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, THEREFORE, A DEVICE ANALYSIS CANNOT BE COMPLETED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED A FOLLOW-UP WILL BE SUBMITTED.
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, DURING INITIAL DRAIN. THERE WERE OPEN CLAMPS ON ANY UNUSED SUPPLY LINES. THE PATIENT WAS CONNECTED EITHER AT THE TIME OF THE ALARM OR OBSERVED AIR. THE PATIENT LINE WAS PROPERLY PRIMED BEFORE CONNECTING. PATIENT EXTENSION LINES WERE USED AND WERE CONNECTED PRIOR TO PRIMING. THE PATIENT DID NOT DISCONNECT ANYTIME PRIOR TO THE ALARM OR OBSERVED AIR. ALL BAGS WERE PROPERLY CONNECTED. THE SUPPLIES WERE NOT DAMAGED BY AN OUTLET PORT CLAMP OR AN ASSIST DEVICE USED TO MAKE THE CONNECTIONS. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE REPORTER. THE HOME PATIENT (HP) WOULD COMPLETE THERAPY WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 134295 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | HOME CHOICE |