FDA Adverse Event Malfunction Summary report: N

ROTABLATOR ROTATIONAL ATHERECTOMY SYSTEM

MDR report key: 3031301 · Received April 2, 2013

Report

Report Number
2134265-2013-02335
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 5, 2013
Report Date
March 5, 2013
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
Product Code
MCX
PMA / PMN Number
P900056
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE BATCH NUMBER IS UNKNOWN AND THE MANUFACTURING RECORDS FOR THE COMPLAINT DEVICE COULD NOT BE REVIEWED. THE MOST PROBABLE ROOT CAUSE IS UNABLE TO BE DETERMINED. (B)(4).

Description of Event or Problem · 1

SAME CASE AS MDR ID 2134265-2013-02271. IT WAS REPORTED THAT DURING PREPARATION FOR A ROTATIONAL ATHERECTOMY TREATMENT PROCEDURE, GUIDE WIRE FRACTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE PROXIMAL RIGHT CORONARY ARTERY. WHILE PREPARING THE 1.25MM ROTABLATOR ROTALINK BURR, THE DEVICE WAS BEING TESTED AT A ROTATIONAL SPEED OF APPROXIMATELY 180RPM WHEN THE ROTABLATOR ROTALINK BURR CAME INTO CONTACT WITH THE ROTAWIRE GUIDE WIRE AND FRACTURED THE GUIDE WIRE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134029 ROTABLATOR ROTATIONAL ATHERECTOMY SYSTEM CATHETER, CORONARY, ATHERECTOMY MCX BOSTON SCIENTIFIC - COSTA RICA (COYOL) UNK509

Patients

Seq Age Sex Outcome Treatment
1