FDA Adverse Event Injury Summary report: N

M2A-T M/H RAD 2HL SHL 41/52MM

MDR report key: 3031222 · Received April 2, 2013

Report

Report Number
0001825034-2013-00827
Event Type
Injury
Date Received
April 2, 2013
Date of Event
March 26, 2013
Report Date
March 7, 2013
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK993438
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, "MATERIAL SENSITIVITY REACTIONS." NUMBER 8 STATES, "DISLOCATION AND SUBLUXATION DUE TO INADEQUATE FIXATION AND IMPROPER POSITIONING. MUSCLE AND FIBROUS TISSUE LAXITY CAN ALSO CONTRIBUTE TO THESE CONDITIONS." NUMBER 14 STATES, "POSTOPERATIVE BONE FRACTURE AND PAIN." THIS REPORT IS NUMBER 3 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2013-00825 / 00827).

Additional Manufacturer Narrative · 1

DIMENSIONAL EVALUATION FOUND COMPONENT TO BE WITHIN APPROPRIATE DESIGN SPECIFICATION. EVALUATION OF THE EXPLANTED DEVICE FOUND THE HEAD AND CUP APPEARED TO SHOW EVIDENCE OF SUBLUXATION OF THE JOINT, SCRATCHING OF THE BEARING SURFACES AND METALLIC TRANSFER MARKS CONSISTENT WITH THE REPORTED DISLOCATIONS. THIS REPORT IS NUMBER 3 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2013-00825-1 / 00827-1).

Description of Event or Problem · 1

IT WAS REPORTED PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2013 DUE TO PAIN, DISCOMFORT, DISLOCATIONS AND A PSEUDOTUMOR. ALL PRODUCT WAS REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134254 M2A-T M/H RAD 2HL SHL 41/52MM PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 602870

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R