FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 3031219 · Received April 2, 2013

Report

Report Number
3015876-2013-00262
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 6, 2013
Report Date
March 6, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K102972
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

PHYSIO REPLACED THE THERAPY CONNECTOR ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED THERAPY CONNECTOR ASSEMBLY AT THE FAILURE ANALYSIS CENTER AND DETERMINED THE CAUSE OF THE FAILURE TO BE DUE TO THE APEX SOCKET MISSING THE INSERT SLEEVE WHICH WOULD NOT ALLOW THE SOCKET TO MAKE CONTACT WITH AN INSERTED PIN.

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED INTERMITTENT FAILURE. PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED EVENT AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE INTERMITTENTLY DETECTS A CONNECTION TO THE HARD PADDLES ASSEMBLY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.