FDA Adverse Event Injury Summary report: N

SELOX JT 45

MDR report key: 3031183 · Received April 2, 2013

Report

Report Number
1028232-2013-00906
Event Type
Injury
Date Received
April 2, 2013
Date of Event
February 28, 2013
Report Date
March 25, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE EVENT AND EXPLANT DATE HAVE BEEN CORRECTED. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE WERE RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.

Description of Event or Problem · 1

THIS SYSTEM WAS REMOVED DUE TO (B)(6). CULTURES OF THE DEVICE AND LEADS SHOWED NO INFECTION. THERE IS NO INDICATION THAT THE SYSTEM WAS REPLACED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134371 SELOX JT 45 PACER LEAD NVN BIOTRONIK SE & CO. KG 346369

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization