FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3031065 · Received April 2, 2013

Report

Report Number
2531779-2013-03439
Event Type
Malfunction
Date Received
April 2, 2013
Report Date
March 5, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/05/2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE DISPLAY SCREEN WAS FOUND TO BE DISCOLORED/FADED. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED AND THE BATTERY CAP THREADS WERE STRIPPED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DISCOLORED/FADED DISPLAY SCREEN. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
135123 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 37 YR