FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMPORO 2

MDR report key: 3031057 · Received March 28, 2013

Report

Report Number
2242352-2013-00296
Event Type
Malfunction
Date Received
March 28, 2013
Date of Event
March 11, 2013
Report Date
March 11, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K101274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE HEMOPRO 2 FAILED TO CUT AND CAUTERIZE. THEY FIRED IT AND LET IT COOL DOWN FOR 30 SECONDS; HOWEVER, IT STILL DID NOT WORK PROPERLY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS. THE HOSP INTENDS TO RETURN THE PRODUCT IN QUESTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
128700 VASOVIEW HEMPORO 2 ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-4000 25069390

Patients

Seq Age Sex Outcome Treatment
1 NA