GYNECARE MORCELLEX* TISSUE MORCELLATOR
Report
- Report Number
- 2210968-2013-03309
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- March 13, 2013
- Report Date
- March 13, 2013
- Manufacturer
- ETHICON, INC.
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.THIS IS ONE OF THREE MEDWATCHES BEING SUBMITTED AS THREE DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2013-03308 AND MEDWATCH 2210968-2013-03310. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.
DATE SENT TO THE FDA: 06/26/2013. CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE BLADE FAILED TO ROTATE DURING THE EVALUATION; IT IS LIKELY THAT THE ACCUMULATION OF BLOOD, BODY FLUIDS, AND TISSUE PREVENTED THE DEVICE FROM CONTINUING TO OPERATE AS INTENDED.
(B)(4). IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE BLADE OF THE DEVICE SLOWED DOWN AND STOPPED SPINNING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 135096 | GYNECARE MORCELLEX* TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC | HET | ETHICON, INC. | NA | MT217316 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |