FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 3030873 · Received April 2, 2013

Report

Report Number
1644487-2013-00891
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
November 7, 2006
Report Date
March 13, 2013
Manufacturer
CYBERONICS
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

WHILE PERFORMING A BATTERY LIFE CALCULATION IT WAS NOTED THAT ON (B)(6) 2006 THE DEVICE WAS INTERROGATED, THEN PROGRAMMED TO 2MA / 30 SF / 500 PW / 30 SECONDS ON TIME / 5 MINUTES OFF TIME. A SYSTEM DIAGNOSTIC RESULTED IN "FAULT" AND CAUSED THE SETTINGS TO CHANGE TO 1/20/500/30/60 MINUTES OFF TIME, PRIOR TO THE PATIENT LEAVING THE VISIT THE DEVICE WAS REPROGRAMMED, HOWEVER NOT EVERY PARAMETER WAS CORRECTED AND THE PATIENT LEFT THE VISIT AT 2.0/30/500/30/60 MINUTES OFF TIME. THEIR OFF TIME WAS CORRECTED ON (B)(6) 2009.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
135338 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS 250 521422

Patients

Seq Age Sex Outcome Treatment
1 50 YR