FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3030822 · Received March 27, 2013

Report

Report Number
2518422-2013-00446
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S OXYGEN BLENDING MODULE BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PT USE.