FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 3030789
·
Received March 27, 2013
Report
- Report Number
- 2518422-2013-00461
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- February 27, 2013
- Report Date
- February 27, 2013
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE INCREASED AIRSTREAM TEMPERATURE (HIGH TEMPERATURE ALARM) APPEARS TO HAVE BEEN CAUSED BY THE AMBIENT CONDITIONS IN WHICH THE DEVICE WAS OPERATING. NO MALFUNCTIONS WERE NOTED.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WAS EMITTING WARM AIR. THE DEVICE WAS NOT IN PT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126107 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC. | 1054260 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |