FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK CR(R) PLUS DEFIBRILLATOR
MDR report key: 3030783
·
Received April 2, 2013
Report
- Report Number
- 3015876-2013-00259
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- March 5, 2013
- Report Date
- March 5, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K033275
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A SHORTED AND BURNED CAPACITOR, DESIGNATOR C76 FROM THE DIGITAL PCB ASSEMBLY. THE SHORTED AND BURNED CAPACITOR CAUSED HIGH CURRENT WHEN POWERED OFF, WHICH DEPLETED THE INTERNAL HLC BATTERIES AND EXTERNAL CHARGE-PAK ASSEMBLY. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, SERVICE WRENCH AND CHARGE-PAK) AND WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.