FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 3030724 · Received March 27, 2013

Report

Report Number
1828100-2013-00304
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE BIOMED IS PLANNING ON REPLACING THE OVER-TEMP THERMOSTAT.

Description of Event or Problem · 1

THE USER FACILITY'S BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE OVER-TEMP 43+ DEGREE THERMOSTAT DID NOT WORK PROPERLY, AS THE COOLER HEATER UNIT WENT TO 46+ DEGREES WHEN TESTING. THE OVER-TEMP LIGHT DID NOT COME ON EITHER. SINCE THE EVENT OCCURRED DURING PREVENTIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.