FDA Adverse Event Injury Summary report: N

VENATECH LP VENA CAVA FILTER

MDR report key: 3030684 · Received March 22, 2013

Report

Report Number
9612452-2013-00001
Event Type
Injury
Date Received
March 22, 2013
Date of Event
January 30, 2013
Report Date
March 21, 2013
Manufacturer
B. BRAUN MEDICAL SAS
Product Code
DTK
PMA / PMN Number
K010485
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW: WE HAVE CHECKED THE MFG FILE OF THIS LOT OF VENA CAVA FILTERS, WHICH COMPLIES WITH OUR SPECS AND DOES NOT PRESENT ANY ABNORMALITY. NO OTHER SIMILAR INCIDENTS WERE DECLARED TO US CONCERNING THIS LOT OF 58 VENA CAVA FILTERS, SOLD SINCE AUGUST 2012. INVESTIGATION RESULTS: THE INVOLVED DEVICE IS NOT AVAILABLE FOR INVESTIGATION NOR THE X-RAY PICTURES, CONSEQUENTLY NO THOROUGH INVESTIGATION CAN BE PERFORMED. CONCLUSION: AS NEITHER THE DEVICE NOT X-RAY PICTURES COPY WERE RETURNED TO US FOR EVAL, WE CANNOT CONCLUDE TO THE REAL CAUSE OF THIS INCIDENT. THIS CASE IS REGISTERED FOR STATISTICAL PURPOSE. THIS DATABASE, WHICH WE ACTIVELY MONITOR AND TREND TO ENSURE WE REACT QUICKLY AND EFFECTIVELY TO THE EXPERIENCES REPORTED FROM THE USE OF OUR PRODUCTS. (B)(4). NO CORRECTIVE ACTION IS ENVISAGED.

Description of Event or Problem · 1

A VENATECH LP VENA CAVA FILTER WAS PLACED IN THE PT ON (B)(6) 2013 IN THE INTERVENTIONAL RADIOLOGY DEPARTMENT DUE TO DVT. ON (B)(6) 2013, THE PT HAD A CHEST CT TO RULE OUT PE. THE CHEST CT SHOWED MULTIPLE BILATERAL PULMONARY EMBOLI WITH THE VENOUS FILTER SEEN WITHIN THE INTRAHEPATIC SEGMENT OF THE INFERIOR VENA CAVA. ON (B)(6) 2013, A CHEST CT SHOWED THE FILTER WAS LARGELY IN THE RIGHT VENTRICLE WITH PARTIAL ATTACHMENT TO THE TRICUSPID VALVE. ON (B)(6) 2013, THE PT HAD SURGERY TO REMOVE THE DISLODGED CAVAL FILTER WITHIN THE RIGHT VENTRICLE. THE PT HAD A NODAL RHYTHM THAT REQUIRED TEMPORARY PACING, AND LATER, A PACEMAKER WAS PLACED ON (B)(6) 2013. THE PT WAS DISCHARGED FROM THE INPATIENT HOSPITAL SETTING TO INPATIENT REHAB ON (B)(6) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
119565 VENATECH LP VENA CAVA FILTER VENA CAVA FILTER DTK B. BRAUN MEDICAL SAS 31335 L198756V

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R