FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3030659 · Received March 27, 2013

Report

Report Number
2518422-2013-00445
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126172 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054096

Patients

Seq Age Sex Outcome Treatment
1