FDA Adverse Event
Malfunction
Summary report: N
VASOVIEW HEMOPRO 2
MDR report key: 3030568
·
Received March 27, 2013
Report
- Report Number
- 2242352-2013-00294
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- March 5, 2013
- Report Date
- March 11, 2013
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE HAS BEEN RETURNED TO THE FACTORY AND IS BEING EVALUATED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE USING THE HEMOPRO 2, IT WAS OBSERVED THAT THE C-RING HAD HEAT RELATED DAMAGE AS IT BROKE IN HALF AFTER REMOVING THE DEVICE FROM THE PATIENT. NOTHING FELL INTO THE PATIENT AND NO INTERVENTION WAS REQUIRED. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126461 | VASOVIEW HEMOPRO 2 | ENDOSCOPIC VESSEL HARVESTING | GEI | MAQUET CARDIOVASCULAR, LLC | VH-4000 | 25069390 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |