FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP** ENDOSCOPIC CLIP APPLIER
MDR report key: 3030518
·
Received April 2, 2013
Report
- Report Number
- 3005075853-2013-01530
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- February 13, 2013
- Report Date
- March 28, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED IN GOOD VISUAL CONDITION. THE JAWS WERE FOUND PROPERLY ALIGNED. IN AN ATTEMPT TO REPLICATE THE INCIDENT REPORTED THE DEVICE WAS FUNCTIONALLY EVALUATED. UPON FIRING OF THE DEVICE, THE REMAINING CLIPS WERE EJECTED AND THEN, IT LOCKED OUT AS INTENDED. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSE THE FOUND CONDITION. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT BEFORE AN UNKNOWN PROCEDURE, WHEN THE STERILE PACKAGE WAS OPENED THE STAPLES DROPPED. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 135006 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | H43P83 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |