FDA Adverse Event Malfunction Summary report: N

MITEK OMNISPAN MENISCAL REPAIR SYSTEM, W/ 12 DEGREE NEEDLE

MDR report key: 3030485 · Received April 2, 2013

Report

Report Number
1221934-2013-00087
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 29, 2013
Report Date
March 29, 2013
Manufacturer
DEPUY MITEK
Product Code
MBI
PMA / PMN Number
K092836
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MITEK IS AT THIS POINT IN TIME IN THE INFORMATION GATHERING MODE. WHEN ALL THAT CAN BE HAD, IS HAD AND THOROUGHLY INVESTIGATED AND EVALUATED, THOSE RESULTS WILL BE THE SUBJECT MATTER IN A FOLLOW-UP REPORT. AWAITING RETURN.

Additional Manufacturer Narrative · 1

THE COMPLAINT DEVICE IS NOT BEING RETURNED, WHICH PRECLUDES CONDUCTING AN EVALUATION; HOWEVER, A BATCH RECORD REVIEW HAS BEEN CONDUCTED TO DETERMINE IF THERE WERE ANY INTERNAL PROCESSING ISSUES WHICH WOULD HAVE CONTRIBUTED TO THE NATURE OF THE PRODUCT COMPLAINT. OUR RESULTS INDICATE THAT THIS BATCH OF PRODUCT WAS PROCESSED WITHOUT INCIDENT AND THEREFORE THERE IS NO INTERNALLY ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM. FURTHER, A REVIEW INTO THE MITEK COMPLAINTS SYSTEM REVEALED NO OTHER SIMILAR COMPLAINTS FOR THIS LOT OF DEVICES THAT WERE RELEASED TO DISTRIBUTION. WE CANNOT TELL ANYTHING FROM THIS; WE CANNOT DISCERN A ROOT OR UNDERLYING CAUSE FOR THE REPORTED FAILURE MODE. AT THIS POINT IN TIME, NO FURTHER ACTION IS WARRANTED. HOWEVER, THIS FILE WILL REMAIN RECEPTIVE TO ANY POTENTIAL FORTHCOMING INFORMATION RECEIVED THAT IS PERTINENT AND GERMANE TO THIS ISSUE. ALSO, MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.

Description of Event or Problem · 1

OUR REP. IS REPORTING TO US THAT DURING A MENISCAL REPAIR WITH THE USE OF OMNISPAN FOR FASTENING, THE SILICONE RETENTION TUBING CAME OFF OF THE INSERTER NEEDLE AND FELL INTO THE PATIENT'S JOINT SPACE. THE TUBING WAS RETRIEVED AND THE PROCEDURE WAS CONCLUDED SUCCESSFULLY WITHOUT FURTHER ISSUE OR HARM TO THE PATIENT

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134055 MITEK OMNISPAN MENISCAL REPAIR SYSTEM, W/ 12 DEGREE NEEDLE MENISCAL FASTENER MBI DEPUY MITEK NA 3662569

Patients

Seq Age Sex Outcome Treatment
1