COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910)

MDR report key: 3030475 · Received April 2, 2013

Report

Report Number
2210968-2013-03271
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 14, 2013
Report Date
March 14, 2013
Manufacturer
ETHICON, INC.
Product Code
GAM
PMA / PMN Number
K032420
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: THE ACTUAL NEEDLE REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. THE DEVICE INFORMATION FOR USE CAUTIONS THE USER THAT "TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT."

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2013 AND SUTURE WAS USED. WHILE SUTURING THE FASCIA, IT WAS NOTED THAT THE NEEDLE TIP BROKE. THE NEEDLE FRAGMENT REMAINED IN THE NEEDLE HOLDER. THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.