ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-03402
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Report Date
- March 6, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 SUBMITTED: 07/23/2013, DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS INTACT WITHOUT DAMAGE. ON TESTING, ALL THE KEYPAD BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND DID NOT REVEAL AND EVIDENCE OF CONTAMINATION UNDER THE BUTTON CONTACTS. INVESTIGATION WAS UNABLE TO DUPLICATE THE COMPLAINT.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE OK KEYPAD BUTTON REQUIRED MULTIPLE BUTTON PRESSES TO ELICIT A RESPONSE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 135108 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |